Strany 1 - Vaisala Remote Display Panel

User's GuideM211691EN-AVaisala Remote Display PanelRDP100

Strany 2

WiringThe wiring instruction below applies to all compatible measurement instrumentsexcept the HMDW110 series (HMDW110 wiring is covered in section Wi

Strany 3 - RDP100 Remote Display Panel

Wiring HMDW110 with RDP100You must always connect the humidity measurement current loop (HUM,terminals 5and 6) to power the transmitter. Connecting t

Strany 4 - Serial Port Settings

HMDW110 Power Supply RequirementsHMDW110 series transmitters are designed for a supply voltage range of10...28VDC. The minimum required voltage dep

Strany 5 - RDP100 Parts

Problem Possible Cause SolutionThe parameter I wantis not selectable.Connected instrumentdoes not provide thedesired parameter.Check the instrument&ap

Strany 6 - Installation

Download manuals from:www.vaisala.com/manualsTechnical support by e-mail:[email protected] information:www.vaisala.com/warrantyVaisala Serv

Strany 7

PUBLISHED BYVaisala OyjStreet address: Vanha Nurmijärventie 21, FI-01670 Vantaa, FinlandMailing address: P.O. Box 26, FI-00421 Helsinki, FinlandPhone:

Strany 8

RDP100 Remote Display PanelOverviewVaisala Remote Display Panel RDP100 is a digital display that shows the livemeasurement of one or two environmental

Strany 9

Serial Port SettingsRDP100 and the measuring instrument communicate on the RS-485 line using aVaisala proprietary protocol. There must be only one RDP

Strany 10

RDP100 Parts1 = Detachable screw terminal, max wire size 1.5 mm2(AWG16).2 = Display element (do not touch).3 = Direction arrow, point up during insta

Strany 11 - Wiring HMDW110 with RDP100

InstallationRequired tools:l Medium size crosshead screwdriver (Pozidriv).l Small slotted screwdriver for screw terminals and DIPswitches.l Drilling

Strany 12 - Troubleshooting

4. Remove the stainless steelfront cover by pulling the bottompart of the cover away from thehousing. You can also insert aslotted screwdriver in the

Strany 13 - Spare Parts

8. Optional:apply sealing paste tothe back frame of the housing(where the housing meets thewall when installed). Follow theinstructions provided with

Strany 14 - *M211691EN*

12. If the DIPswitches on theRDP100 are in off position (newRDP100), the RDP100 promptsyou to select 1-2 parameters fordisplay. Use the DIPswitcheso