Vaisala HMP243 Uživatelská příručka Strana 48

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 130
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 47
HMP240
User's Guide M210300en
46
Insert the sensor head into the measurement hole of the
LiCl chamber.
Wait at least 10 minutes for the reading to stabilize.
Give command CRH <cr>, enter the first point value and
press <cr>.
>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
If you want to see how the sensor stabilizes to the humid-
ity in the calibrator, enter c <cr>:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
Insert the sensor head into the measurement hole of the
NaCl chamber.
Wait at least 20 minutes for the reading to stabilize.
Press any key, enter the second point value and press
<cr>.
RH : xx.x Ref2 ? yy.y <cr>
The stabilization of the sensor can be monitored by
entering c<cr>.
6.1.2.2 With display / keypad commands
Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.
Disconnect the security lock jumper!
Insert the sensor head into the measurement hole of the
LiCl chamber.
Wait at least 20 minutes for the reading to stabilize.
Select Cali in the main menu and then RH cal; select Not
changed and then two-point calibration RH 2 point cal.
Change the first point reading with the arrow keys to cor-
respond to the reference humidity and press ENT; pressing
an arrow once changes the reading by 0.05 %RH.
Zobrazit stránku 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 129 130

Komentáře k této Příručce

Žádné komentáře