Error Indication in Transmitters without
Display
Error codes are cabled by blinking a message with short flashes of the
LED lights.
LEDs are situated on the cover of the transmitter. The green OPERATION
LED is blinking when a non-critical error occurs (no effect on relays). The
red LEVEL 1 and LEVEL 2 LEDs are blinking when a fatal error occurs
(switches the relays OFF).
NOTE
The red level LED is continuously lit when the predefined CO
2
concentration is exceeded.
Error Messages
Table 6 Error Messages
Error
Message
Probable Cause Action Severity
Er 01 Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 02 Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 03 Main board memory
problem
Contact Vaisala Service
Center
Fatal
Er 04 Probe contact failure Contact Vaisala Service
Center
Fatal
Er 05 Probe contact failure Contact Vaisala Service
Center
Fatal
Er 06 Probe contact failure Contact Vaisala Service
Center
Fatal
Er 08 Incompatible probe Contact Vaisala Service
Center
Fatal
Er 10 IR-source failure Contact Vaisala Service
Center
Fatal
Er 11 IR-source failure Contact Vaisala Service
Center
Fatal
Er 12 Sensor failure Contact Vaisala Service
Center
Fatal
Er 13 Signal error None Non-
critical
Er 14 Measurement out of
range
Check environment and
recalibrate if necessary
Non-
critical