Vaisala DRYCAP DMT340 SERIES Uživatelská příručka Strana 161

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 191
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 160
Chapter 7 ___________________________________________________ Calibration and Adjustment
VAISALA _______________________________________________________________________ 159
0706-024
Figure 87 Following Stabilization
5. Purge will be performed next. Follow the purge progress on the
display.
6. Let the sensor stabilize for at least 1 hour after the purge before
proceeding with the adjustment. Press READY when stabilized.
0706-025
Figure 88 Proceeding with T
d/f
Adjustment
7. Enter the actual reference frostpoint temperature.
8. The transmitter performs T
d/f
adjustment. This may take up to one
minute.
0706-026
Figure 89 Completing T
d/f
Adjustment
9. After adjustment wait for sensor to cool down which will take 2
minutes at the maximum. The adjustment is now completed.
NOTE
Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.
Zobrazit stránku 160
1 2 ... 156 157 158 159 160 161 162 163 164 165 166 ... 190 191

Komentáře k této Příručce

Žádné komentáře