Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 41
NOTE
When the DMT340 is installed in a process with a pressure differing
from the selected operating pressure at the time of ordering, please enter
the pressure value of the process into the transmitter memory, see
section Pressure Compensation Setting on page 101. Use the serial line
commands XPRES and PRES or the display/keypad. The pressure set
buttons on the motherboard inside the transmitter can also be used to set
pressure compensation.
DMT347 Small Pressure-tight Probe
The DMT347 is ideal for tight spaces with threaded connection. The
small probe is installed using the threaded fitting bodies, see below.
Pressure-Tight Swagelok Installation Kits
for DMT347
Swagelok installation kit for the dewpoint probe includes Swagelok
connector with ISO1/2" thread (Vaisala order code: SWG12ISO12),
ISO3/8" thread (Vaisala order code: SWG12ISO38) or NPT1/2" thread
(Vaisala order code: SWG12NPT12).
0503-042
Figure 24 DMT347 Probe with Swagelok Installation Kit