Technical Data
Measured Variables DMT345
DEWPOINT DMT345
Sensor Vaisala DRYCAP
®
180S
Measurement range -40 ... +100 °C (-40 ... +212 °F) Td
Accuracy ±2°C (±3.6 °F) Td
See the accuracy graph below
DMT346, Reliable in Very Hot Processes
When process temperatures range between 140 °C to
350 °C, the DMT346 provides the best measurement
performance.
The DMT346 comes with a cooling set as a standard
feature. The cooling effect may be regulated by adding the
cooling ns, or removing them from the set for the best
measurement performance.
The cooling system operates without moving parts,
additional power or cooling utilities, thereby eliminating
the risk of sensor damage due to a mechanical cooling
failure.
Additionally, sensor warming minimizes the risk of
condensing on the sensor. In low humidity the combination
of auto-calibration and DRYCAP
®
ensures accurate
measurement.
Graphical Display
The DMT345 and DMT346 transmitters can be ordered
with a large numerical and graphical display, which allows
the user to clearly monitor operational data, measurement
trends and up to one-year measurement history.
The optional data logger with real-time clock makes it
possible to generate more than four years of the measured
history and zoom in on any desired time or time frame.
Versatile Outputs and (Wireless) Data
Collection
The transmitter can be connected to a network with an
optional (W)LAN interface, which enables a (wireless)
Ethernet connection.
For serial interface also the USB connection, RS232 and
RS485 can be used. Additionally an alarm relay option is
available.
The transmitter can have up to three analog outputs.
Galvanic isolation of supply power and analog outputs
are also offered. The recorded measurement data can be
viewed on the display or transferred to a PC with Microsoft
Windows
®
software.
DMT345-346 is also capable in applying the MODBUS
communication protocol and together with an appropriate
connection option provides either MODBUS RTU (RS485)
or MODBUS TCP/IP (Ethernet) communication.
Units are delivered installation-ready.
TEMPERATURE DMT345
Measurement range 0 ... +180 °C (+32 ... +356°F)
with sensor warming upper range limited by humidity
(at 80% RH warming is switched on and T
reading not actual process Temperature)
Accuracy ±0.4 °C at 100 °C
Temperature sensor Pt 100 IEC 751 1/3 class B
RELATIVE HUMIDITY DMT345
Measurement range 0 ... 100 % RH
with sensor warming 0 ... 80 % RH
Accuracy
below 10 % RH ±10 % of reading
above 10 % RH ±1,5% RH + 1,5 % of reading
MIXING RATIO DMT345
Measurement range (typical) 0 ... 1000 g/kg (0 ... 7000 gr/lbs)
Accuracy (typical) ±12% of reading
Response time 63 % [90 %] flow rate 1 l/min and 1 bar pressure
from dry to wet 5s [10 s]
from wet to dry including auto-calibration 45s [5 min]
Dewpoint accuracy vs. measurement conditions