>SERI O <cr> changing parity only
4800 O 7 1 HDX
>SERI 600 N 8 1 F <cr> changing all parameters
600 N 8 1 FDX
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.
5.4.2.6 Selecting the output units
UNIT x <cr>
x = m(etric units)
n(on-metric units)
metric non-metric
RH %RH %RH
T °C °F
Td °C °F
dT °C °F
a g/m
3
gr/ft
3
x g/kg gr/lb
Tw °C °F
5.4.2.7 Setting the averaging time
Disconnect the security lock jumper!
FILT nnnn <cr>
nnn = averaging time (0 - 1024 seconds)
This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.