When replacing the filter, note the following:
- Touching the filter with your fingers will also deposit oil and dirt on
the filter, which may affect the measurement. Always handle the filter
with gloved hands.
- When installing the transmitter after a filter change, inspect the
sealing ring first. Use a new one if the old ring is visibly damaged.
1102-003
Figure 24 Opening the Filter
1102-004
Figure 25 Filter Structure
where
1 = Sealing ring (replace if damaged)
2 = Vaisala HUMICAP® 180R sensor and Pt1000 temperature
sensor
3 = Spring washer
4 = Tube filter
Replace the filter as follows:
1. Turn the filter counterclockwise until it is loose.
2. Pull the filter straight out carefully; do not damage the sensors, and
do not lose the spring washer.