Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 35
Stop Continuous Outputting
Use the S command to stop the continuous measurement output:
S<cr>
Set/Show Output Interval
Use the INTV command to show or set the output interval of the serial
line measurement messages (applies when R command or RUN mode is
used). The shortest output interval is one second. This command has no
effect on the operation of the analog outputs.
INTV [n xxx]<cr>
where
n
= Time interval in range 1 ... 255
xxx
= Time unit = "S", "MIN", or "H"
Examples:
>intv
Value : 1
Unit : S
>
>intv 3 min
Value : 3
Unit : MIN
>
Output a Reading Once
Use the SEND command to output a single measurement reading from
the transmitter.
SEND [X]<cr>
Example:
>send
RH= 25.12 % T= 24.91 'C
>
With X parameter the transmitter will output RH + T value in predefined
format regardless of what quantities are selected.