User's Guide ______________________________________________________________________
50 __________________________________________________________________ M210566EN-J
Installation
1. Disconnect the power and open the transmitter cover.
2. Remove the protective plug from the cable gland and thread the
wires. In case the power supply module is installed in the factory,
continue with the step 5.
3. Attach the power module to the bottom of the housing with four
screws. See the position Figure 2 on page 22.
4. Connect the wires from the terminals of the power supply module
marked with + and - to the terminals POWER 10 ... 35 V 24V on
the motherboard of the transmitter.
5. Connect the AC mains voltage wires to the power supply module
terminals marked with N and L.
6. Attach the grounding wire to the grounding terminal on the right-
hand side of the transmitter.
7. Connect the power. The LED on the cover of the transmitter is lit
continuously during normal operation.
WARNING
Do not detach the power supply module from the transmitter when the
power is on.
WARNING
Do not connect the mains power to power supply module when it is not
installed in the transmitter.
WARNING
Always connect the protective ground terminal.