Chapter 4 ________________________________________________________________ Operating
VAISALA _______________________________________________________________________ 17
Operating Commands
.P Outputting One Single Message
.P<cr>
The command outputs the latest multi-point corrected pressure reading
with two decimals.
.P<cr> (command invisible)
1012.99
.BP Continuous Outputting
.BP<cr>
The command outputs the MPC corrected pressure reading
continuously. The output interval depends on the measurement rate
and the averaging. The outputting is stopped by pressing enter key
(<cr>).
.P<cr> (command invisible)
997.99
998.01
998.01
...
998.02
<cr> (command invisible)
.UNIT Selecting a Pressure Unit
.UNIT.x<cr>
x = 0 (hPa) 6 (kPa)
1 (mbar) 7 (Pa)
2 (inHg) 8 (mmH
2
O)
3 (psia) 9 (inH
2
O)
4 (torr) 10 (bar)
5 (mmHg
As default, the unit is 0.
NOTE
To activate the new setting, please remember to reset the barometer
with the command .
RESET<cr>