Vaisala PTU200 Series Uživatelská příručka Strana 47

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 64
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 46
CHAPTER 7____________________________________________SELF-TESTING AND PROBLEM HANDLING
VAISALA _________________________________________________________________________ 43
CHAPTER 7 SELF-TESTING AND
PROBLEM HANDLING
Returning the serial communication
parameters
To restore factory software and hardware settings, set the dip switch
S1 on the CPU board as in FIGURE 7-1. Other switches except the SW8
must be OFF. In addition, the serial interface has to be set for basic RS
232C operation according to FIGURE 7-1. With these selections, one
single transmitter can be operated through the RS 232C serial
interface using factory serial bus settings as in TABLE 7-1. Returning
the SW8 to OFF will restore the original, application specific software
settings.
ON
OFF
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
S
1
TXD
RXD
X15
RX/RXD
TX
TXD
FIGURE 7-1 Factory settings for the dip switch S1 and Basic RS
232C settings
TABLE 7-1 Serial interface factory settings
Baud rate 9600
Parity even
Data bits 7
Stop bits 1
Duplex full duplex
Zobrazit stránku 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 63 64

Komentáře k této Příručce

Žádné komentáře