Vaisala HUMICAP HMT310 Uživatelská příručka Strana 70

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 105
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 69
User's Guide _______________________________________________________________________
70 ___________________________________________________________________ M210619EN-E
Calibration and Adjustment Commands
Revert to Factory Calibration
Syntax: LI<cr>
This command reverts only the CRH calibration (see calibration
instructions, starting on page 73).
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Issue the LI command and enter value 0 for offset and value 1 for
gain.
3. Reset the transmitter by issuing the RESET command. The
transmitter returns to normal mode.
Example:
>li
RH offset : -0.6000000 ? 0
RH gain : 1.00000000 ? 1
T offset : 0.00000000 ? 0
T gain : 0.40000000 ? 1
>
View Current Offset and Gain Adjustment
View the currently applied standard user adjustment offset and gain using
the L command.
Syntax: L<cr>
Example:
>l
RH offset : 0.00000000
RH gain : 1.00000000
T offset : 0.00000000
T gain : 1.00000000
>
Zobrazit stránku 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 104 105

Komentáře k této Příručce

Žádné komentáře