Vaisala PTU300 Uživatelská příručka Strana 205

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 217
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 204
Appendix B _______________________________________________________ Calculation Formulas
VAISALA _______________________________________________________________________ 203
APPENDIX B
CALCULATION FORMULAS
This Appendix contains the formulas used for the calculated output
quantities.
The PTU300 series transmitters measure relative humidity and
temperature. From these values dewpoint, mixing ratio, absolute
humidity and enthalpy in normal pressure are calculated using the
following equations:
Dewpoint:
1
log
=
A
Pw
m
T
T
n
d
(1)
P
w
is the water vapor pressure. The parameters A, m, and T
n
depend on
temperature according to the following table:
t
A
m
Tn
<0 °C
1)
6.1134
9.7911
273.47
0 ... 50 °C
6.1078
7.5000
237.3
50 ... 100 °C
5.9987
7.3313
229.1
100 ... 150 °C
5.8493
7.2756
225.0
150 ... 180 °C
6.2301
7.3033
230.0
1)
Used for frostpoint calculation if the dewpoint is negative
Zobrazit stránku 204
1 2 ... 200 201 202 203 204 205 206 207 208 209 210 ... 216 217

Komentáře k této Příručce

Žádné komentáře