Configuring Serial Line Operation
Set Serial Line Operating Mode
Use the SMODE command to set the start-up operating mode of the
serial line.
SMODE [mode]<cr>
where
mode
=
STOP, RUN, or POLL.
Table 11 Selection of Output Modes
Mode
Measurement Output
Available Commands
STOP
Only with the SEND command.
All (default mode).
RUN
Automatic output.
Only command S.
POLL
Only with the SEND [addr]
command.
SEND [addr] and OPEN [addr].
Other commands available after
opening a line to the transmitter
using the OPEN command.
Use with RS-485 buses where
multiple transmitters can share
the same line.
Selected output mode will be activated at next reset or power up.
Example:
smode poll
Serial mode : POLL
Set Serial Line Settings
Use the SERI command to set the serial line settings. The new settings
will be taken into use when the transmitter is reset or powered up.
NOTE
If you have altered the DMT143 serial line settings and start using the
MI70
handheld indicator, always switch off the other power supply
(disconnect “Analog” cable if in use) before powering up the MI70.
For
more information on using the MI70 indicator with DMT143, see section
Operation with DM70 Hand
-Held Dewpoint Meter on page 34.