
User's Guide _______________________________________________________________________
104 __________________________________________________________________ M210762EN-G
Data Filtering
The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 29 Filtering Levels
Standard filtering, enabled by default.
Approximately 13 s moving average.
Extended filtering. Approximately 1 min average by
default, but configurable on the serial line.
Use display/keypad to set the filtering level.
1. Open the Main Menu by pressing any of the ▲▼◄► arrow
buttons.
2. Select Measuring by pressing the ► arrow button.
3. Select Filtering and press CHANGE to confirm your selection.
4. Select Off/Standard/Extended and press SELECT to confirm
your selection.
5. Press EXIT to return to the basic display.
Komentáře k této Příručce