Chapter 15 _____________________________________________________ Technical Specifications
VAISALA _______________________________________________________________________ 53
Data logging capacity 900...2700 real time data points,
Logging interval 1 s to 12 h
Logging duration 1 min ... memory full
Alarm Audible alarm function
Housing classification IP54
Weight 400 g
Housing material ABS/PC-blend
Battery pack
Operation times in
continuous use 48 h typical at +20 °C (68 ° F)
data logging use up to 30 days
Power consumption during charge 10W max
Charging time 4 hour
General about the MM70 Hand held humidity and temperature meter
Storage temperature -40 ...+70 °C
Storage humidity range 0...100 %Non Condensing
Electromagnetic compatibility
Complies with the following standard:
EN 61326-1:1997 +Am 1:1998, Electrical equipment for measurement, control and
laboratory use - EMC requirements: Portable test and measurement equipment.