Vaisala HMP60 Uživatelská příručka Strana 48

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 73
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 47
User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M211060EN-F
Clear Adjustment of T Measurement
Use the CTCLR command to clear the adjustment of temperature
measurement that has been done using the CT command.
CTCLR<cr>
Example:
ctclr
OK
View User Adjustment Parameters
Use the L command to view the current user adjustment parameters. This
command is useful for checking the currently applied customer
calibration.
L<cr>
The output values are as follows:
- Cp offset and gain: capacitance, calibrated using the CRH command
- T offset and gain: calibrated using the CT command
Example (shows default values, no user calibration done):
l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Zobrazit stránku 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 72 73

Komentáře k této Příručce

Žádné komentáře