User’s Guide ______________________________________________________________________
28 ___________________________________________________________________ M010056EN-J
0505-274
Figure 11 Cleaning the Tightening Cone
The following numbers refer to Figure 11 on page 28:
1 = Fitting screw
2 = Sealing washer
3 = Tightening cone
4 = Clean cotton stick
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
NOTE
When HMP364 is installed in a process with a pressure differing from
normal atmospheric pressure, please enter the pressure value of the
process (in bar
a
) into the transmitter memory, see sections PRES Setting
Ambient Pressure for Calculations on page 56 and Setting the Pressure for
Calculations on page 46.