HMP368 for Pressurized Pipelines or
Moisture in Liquids
Due to the sliding fit the HMP368 is easy to install into and remove from
a pressurized process. The probe is especially suitable for measurements
inside pipelines. See section Installing the HMP368 Probe Through a Ball
Valve Assembly on page 32.
0603-013
Figure 13 HMP368 Probe
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
NOTE
When HMP368 is installed in a process with a pressure differing from
normal atmospheric pressure, please enter the pressure value of the
process (in bar
a
) into the transmitter memory. See sections PRES Setting
Ambient Pressure for Calculations on page 56 and Setting the Pressure for
Calculations on page 46.
The following numbers refer to Figure 13 on page 30:
1 = Clasp nut, 24 mm hex nut
2 = Fitting body, 27 mm hex nut