Vaisala HUMICAP HMW90 SERIES Specifikace Strana 31

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 74
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 30
525
HUMIDITY
HUMIDITY
9
WE MAKE IT EASY.
kele.com 888-397-5353 USA
March 2014
ROOM, DUCT, AND OSA 2% HUMIDITY TRANSMITTERS
SRH12 SERIES
Accuracy 2% RH (0-100% RH) @ 73°F (23°C)
Drift <1% RH/yr @ 73°F (23°C), 50% RH
Supply Voltage 12-30 VDC
Signal Output 4-20 mA, 2-wire
Maximum Output
Impedance 700Ω @ 24 VDC Ω = (supply - 10) /
0.02
Sensing Technology Replaceable capacitive
Sensor Protection Porous Polyethylene
Protection Reverse polarity protection
Operating Humidity 0 to100%
Operating Temperature -40° to 140°F (-40° to 60°C)
Enclosure
Room ABS plastic (flammability, 94V-0)
Duct Polycarbonate (flammability, 94V-0),
NEMA 4 (IP65),
Wiring Terminations Pluggable terminal block
Dimensions
Room 4.7"H x 2.8"W x 1.0"D
(12 x 7 x 2.5 cm)
Duct 5.1"L x 2.7"W x 2.4"D
(13 x 6.8 x 7 cm)
8.5" (21.6 cm) probe length 1/2"
probe Ø
OSA 6.1"L x 4"W x 12.5"D
(15.5 x 10 x 32 cm)
Approvals CE, RoHS
Weight
Room 0.5 lb (0.23 kg)
Duct 1.3 lb (0.59 kg)
OSA 2.6 lb (1.2 kg)
Warranty 1 year (sensor) 3 years
(electronics only)
DESCRIPTION
Designed for the HVAC control system, the Setra SRH12
Series of humidity transmitters offer reliable measurement of
RH levels from 0 to 100% with an accuracy of 2%. Available in
wall, duct, and outside air configurations, the SRH12 Series
is easy to install and service if necessary. The transmitters
utilize a field replaceable sensor tip that allows the end user
to replace sensors on-site. This eliminates factory calibration
and downtime.
FEATURES
±2% RH accuracy
No recalibration required
Replaceable sensor
Capacitive sensor technology
Reverse polarity protection
Shield
RTN/COM
+EXC
6
5
4
3
Ground connection
(not used)
(not used)
(not used)
2 RH (4 to 20 mA) output
1 Vin
WIRING
APPLICATION
HVAC/R building controls
Indoor air quality
Laboratories
Antiquities preservation
Swimming pools
Clean rooms
SPECIFICATIONS
Outside Air
Duct
Wall
Zobrazit stránku 30