Technical data
Ref. B210823EN-A ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Performance
Measurement range 0 ... 100 % O
2
Accuracy (including noise, linearity and
repeatability) ±0.2 % O
2
Temperature range over T range ±2 % of reading
Stability ±1 % of reading/yr
Zero drift ±0.1 % O
2/
yr
Response time of measurement 3 s
Diffusion limited response in still air T
63
/T
90
without filters 10 s / 20 s
with stainless steel mesh filter 10 s / 25 s
with stainless steel mesh filter and
PTFE filters 30 s / 70 s
Pressure dependence without pressure
compensation
0.8 ... 1.2 bar -2 % of reading
1.2 ... 1.4 bar -5 % of reading
Accuracy of pressure compensation ±0.25 % of reading
Background gas effects for CO
2
and H
2
O uncompensated
< 1% of reading for <6 vol -% CO
2
< 1% of reading for gas dew point < 30 ˚C
Accuracy of background gas compensation
0 ... 50 vol-% CO
2
±-0.5 % of reading
0 ... 300 g/m
3
H
2
O (T
d
=80 ˚C) ±1 % of reading
Classification
OMT364 CLASSIFICATIONS FOR THE TRANSMITTER
USA and Canada FM, Non-incendive for Class I,
Division 2, Groups A, B, C, and D.
T
a
= -20 ˚C to +60 ˚C
OMT364 CLASSIFICATIONS FOR THE PROBE
USA and Canada FM, Suitable for Class I, Division 1,
Groups A, B, C, and D.
T
a
= -20 ˚C to +80 ˚C
Inputs and outputs
Power supply
input range 11 ... 36 VDC
Power consumption
maximum 6W
typical 3W
Analog output 0/4 ... 20 mA sourcing
maximum load 500 
Serial output (2-wire, non-isolated) RS-485
Alarm/control relay 30 VAC/60 VDC
Operating environment
Operating temperature range for
probe (in-line installation) -20 ... +80 ˚C
electronics (housing) (Kalrez® seals)
-20 ... +60 ˚C
transmitter (ambient gas
measurement) -20 ... +60 ˚C
Storage temperature range -55 ... +80 ˚C
Operating pressure range 0.8 ... 1.4 bar
Maximum pressure range for probe up to 10 bar
Compliance
IEC(EN)-61326 Electrical equipment for measurement,
control and laboratory use. EMC requirements.
Laser safety Class 1 laser product
Mechanics
Housing material G-AlSI10 Mg (DIN 1725)
Probe material AISI 316
Housing classification IP66
Weight, transmitter and probe 2.2 kg
Weight, sample cell 2.6 kg
Mounting flance diameter 97 mm
Cable bushing Cable gland M20 x 1.5
Stainless steel mesh filter openings ø 310 m
Wetted materials
AISI 316, Kalrez
®
, PTFE (optional),
SiN, MgF
2
Options and accessories
Hydrophobic PTFE filter, pore size 8 m with
Kalrez
®
O-rings 217056
1/2” NPT conduit fitting 217917
Sample cell with wall-mounting bracket
gas fittings for O.D. 6 mm tube
volume 260 m
3
T63 response time with 1 l/min
sample flow and mesh filter, 18 s
with Kalrez
®
O-rings 216619
OMT364 Oxygen Transmitter for
Hazardous Areas
The Vaisala SPECTRACAP
®
Oxygen
Transmitter OMT364 is designed
for oxygen monitoring in moist and
aggressive process gases.
Laser based measurement
The OMT364 incorporates a
compact tunable diode laser (TDL)
gas spectrometer built in the
measurement probe. This optical
measurement technology is well
known for unmatched stability
and robustness. The patented
SPECTRACAP
®
sensor offers uniquely
TDL technology in a compact probe
suitable for eld use.
FM Approved OMT364 for
Hazardous Areas
The non-incendive OMT364
transmitter is permitted for use in
Class I, Division 2 of the NEC
®
500
area classication according to FM
Approvals SM . The probe of the
The Vaisala SPECTRACAP
®
Oxygen Transmitter OMT364 for hazardous areas.
instrument can be installed in areas
that are classied as Class I, Division
1 or 2.
Direct installation
In many applications, the OMT364
can be ange-mounted directly into
a process. No sampling or sample
conditioning equipment is needed.
This feature provides real time
measurement data without sampling
or sample switching delays.
Sampling cell installation
An optional sampling cell is
available for processes with high
temperatures, elevated pressures
or extremely difcult mechanical
conditions. Due to the robustness of
the SPECTRACAP
®
sensor and its low
sensitivity to gas ow and pressure
variations, the sampling system is
simple and it can be installed near
the sampling point.
Features/Benefits
Minimizes need for sample
conditioning equipment
In-situ probe or sampling cell
Tolerates aggressive chemicals
Tolerates excessive amounts of
moisture even in liquid form
Low maintenance
Calibration interval 12 months
+"03.!ŏ* ŏ'!5, ŏ1/!.ŏ
interfaces and an LCD display
Example applications:
fermentation, composting
process monitoring, inerting
processes and flare gas
recovery