153
Technical Data
Pressure
Performance
BAROMETRIC PRESSURE
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa
Accuracy 500 ... 1100 hPa 500 ... 1100 hPa 50 ... 1100 hPa
CLASS A CLASS B
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa
Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Calibration
uncertainty**
±0.07 hPa ±0.15 hPa ±0.20 hPa
Accuracy at +20 °C*** ±0.10 hPa ±0.20 hPa ±0.30 hPa
Temperature
dependence****
±0.1 hPa ±0.1 hPa ±0.3 hPa
Total accuracy
(-40 ... +60 °C/
-40 ... +140 °F)
±0.15 hPa ±0.25 hPa ±0.45 hPa
Long-term stability/year ±0.1 hPa ±0.1 hPa ±0.2 hPa
Response time (100 % response)
one sensor 2 s
•
1 s
•
1 s
•
Pressure units
hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr,
psia
* Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis error or repeatability error and calibration.
** Defined as ±2 standard deviation limits of accuracy in the working
standard including NIST traceability.
*** Defined as the root sum of the squares (RSS) of endpoint non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room
temperature.
**** Defined as ±2 standard deviation limits of temperature dependence over
the operating temperature range.
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity,
hysteresis, and repeatability at
+15 ... +25 °C ±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C ±(1.0 + 0.008 x reading) %RH
at -40 ... +60 °C ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C )
(Defined as ±2 standard deviation
limits. Small variations possible,
see also calibration certificate.)
± 0.6 %RH (0 ... 40 %RH)
± 1.0 %RH (40 ... 97 %RH)
Sensor
for typical applications Vaisala HUMICAP
®
180 or 180R*
for applications with chemical
purge/warmed probe Vaisala HUMICAP
®
180C or 180RC*
Response time (90%) at +20 °C (+68 °F) in still air
with grid filter 8 s / 17 s*
with grid + steel netting filter 20 s / 50 s*
with sintered filter 40 s / 60 s*
* with HUMICAP
®
180R or 180RC sensor
TEMPERATURE
Measurement range, all probes -40 ... +60 °C (-40 ... +140 °F)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.4 °F)
Temperature units °C, °F
Temperature sensor Pt100 RTD Class F0.1 IEC 60751
ACCURACY OVER TEMPERATURE RANGE
Operating Environment
Operating temperature -40 ... +60 °C (-40 ... +140 °F)
with optional display 0 ... +60 °C (+32 ... +140 °F)
Humidity range non-condensing
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Industrial Environment
Note: Transmitter with display test impedance of 40 ohm
is used in IEC61000-4-5 (Surge immunity)
Δ
°C
°C
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
0
0-40 -20 20 40 60
The display also shows the WMO pressure
trend ∆P 3h and tendency of 0 ... 8.
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC ±20%
with optional power supply module 100 ... 240 VAC, 50/60 Hz
Power consumption at +20 °C (U
in
24 VDC)
RS232 max. 28 mA
U
out
3 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 33 mA
I
out
3 x 0 ... 20 mA max. 63 mA
display and backlight +20 mA
during chemical purge max. +110 mA
during probe heating +120 mA
Settling time at power-up (one sensor)
class A 4 s
class B 3 s
External loads
current outputs R
L
< 500 ohm
0 ... 1 V output R
L
> 2 kohm
0 ... 5 V and 0 ... 10 V outputs R
L
> 10 kohm