Pressure Compensation Settings
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the
process pressure is taken into consideration.
Note that conversions from mmHg and inHg are defined at 0°C and
for mmH
2
O and inH
2
O at 4°C.
NOTE
Pressure compensation is intended to be used in normal air only.
When measuring in other gases, please contact Vaisala for further
information.
Using Display/Keypad
Use display/keypad to set the pressure compensation. To select the
pressure unit using display/keypad see section Changing Quantities
and Units on page 73.
1. Press any of the arrow keys to open the MAIN MENU.
2. Select Measuring and press the right arrow key.
3. Select Pressure compensation and press the right arrow key.
4. Use the up/down arrow keys to select either Fixed: 1013.25hPa
or Measured P for pressure compensation.
5. Selecting Measured P: Press the SELECT key and then exit
the menu.
6. Selecting Fixed: 1013.25hPa: Press the SELECT key, and then
SET. You can move from a digit to another one by pressing the
left/right arrow keys. To change the unit, press the up/down
arrow keys.
7. Press the OK key and then exit the menu.
Using Serial Line
PRES and XPRES
Command XPRES should be used if the value is changed frequently.
Its value is not retained at reset, and when set to 0, last value set with
PRES is used instead. Use the serial line and do the following:
PRES [aaaa.a]