Chapter 9 _____________________________________________________________ Other functions
VAISALA _______________________________________________________________________ 33
6. Select Quantity, press SELECT. (One quantity can be chosen at a
time. Only active quantities can be selected. To change the
quantities, go to MENU Display Quantities and units).
7. Select 0.0 V to set the value for the 0.0V output signal, press SET.
(If the analog output is ON, take it OFF) Set the low value by using
arrow buttons. Press +/- button to choose the sign of the value.
Press OK to make the setting.
8. Select 1.0V to set the value for the 1.0V output signal, press SET.
Set the high value by using arrow buttons. Press +/- button to
choose the sign of the value. Press OK to make the setting.
9. Select Analog output on/off, press ON to activate the analog output and
return back to the basic display.
10. A wave picture sign appears on the upper left corner of the basic
display.
11. To deactivate the analog output function, go to
MENU Functions Analog outputAnalog output on/off and press OFF.