USER'S GUIDE____________________________________________________________________
106 _________________________________________________________________ M210762EN-D
3. Select Clear graph memories by pressing the CLEAR button.
Press the YES button to confirm the selection.
CAUTION
This function clears the entire data history of the transmitter, including
all graphs and the content of the optional data logger module.
UNDELETE
Similarly to the DELETE command, the UNDELETE command is used
without any arguments. It will recover all deleted data that has not been
overwritten yet.
Analog Output Settings
The analog outputs are set in the factory according to the order form. In
case you want to change the settings, follow these instructions.
Changing Output Mode and Range
Both output channels have their own dip switch module with 8 switches,
see the position in Figure 2 on page 23 (dip switches for analog output
settings).
1. Select the current/voltage output, switch ON either of the switches,
1 or 2.
2. Select the range, switch ON one of the switches from 3 to 7.