Vaisala DRYCAP DMT345 Uživatelská příručka Strana 153

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 157
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 152
Appendix B ______________________________________________________ Calculation Formulas
VAISALA ______________________________________________________________________ 153
APPENDIX B
CALCULATION FORMULAS
This Appendix contains the formulas used for the calculated output
quantities.
The DMT346 series transmitters measure relative humidity and
temperature. From these values dewpoint, mixing ratio, absolute
humidity and enthalpy in normal pressure are calculated using the
following equations:
Dewpoint:
1
log
A
Pw
m
T
T
n
d
(1)
P
w
is the water vapor pressure.The parameters A, m, and T
n
depend on
temperature according to the following table:
t A m T
n
<0 °C
1)
6.1134 9.7911 273.47
0 ... 50 °C 6.1078 7.5000 237.3
50 ... 100 °C 5.9987 7.3313 229.1
100 ... 150 °C 5.8493 7.2756 225.0
150 ... 180 °C 6.2301 7.3033 230.0
1) Used for frostpoint calculation if the dewpoint is negative
Mixing ratio:
w
w
Pp
P
x
-
99.621
(2)
Zobrazit stránku 152
1 2 ... 148 149 150 151 152 153 154 155 156 157

Komentáře k této Příručce

Žádné komentáře