Low maintenance
A stainless steel mesh  lter and an
optional porous PTFE  lter protect
the OMT355 probe optics from dust
and dirt. An intelligent measurement
algorithm further minimizes
contamination effects and provides
a maintenance alert well before
the measurement performance is
affected.
Long calibration interval
Calibration or  eld checking of
the instrument can be done either
with ambient air or with zero/span
gases injected through an optional
calibration gas connector. The
calibration interval of the OMT355 is
12 months.
User friendly interfaces
For calibration and con guration, the
OMT355 features both software and
keypad user interfaces and an LCD
display.
TDL Technology
The Tunable Diode Laser
(TDL) technology used in the
SPECTRACAP
®
sensor is one of the
most powerful technologies in gas
measurement on the market today.
The technology is based on
measuring the attenuation of a beam
of laser light in the sample gas. For
oxygen sensing the laser is tuned to
a wavelength which is characteristic
and unique for the oxygen molecule.
Therefore, the measured attenuation
is an extremely selective measure for
the amount of oxygen on the path of
the laser beam.
Technical data
Performance
Measurement range 0 ... 100 % O
2
Accuracy
(including noise, linearity, and repeatability)
±0.2 % O
2
Temperature dependence over T range ±2 % of reading
Stability ±1 % of reading/yr
Zero drift ±0.1 % O
2
/yr
Response time of measurement 3 s
Diffusion limited response in still air T
63
/T
90
without fi lters 10 s / 20 s
with stainless steel mesh fi lter 10 s / 25 s
with stainless steel mesh and PTFE fi lters 30 s / 70 s
Pressure dependence without pressure compensation
0.8 ... 1.2 bar -2 % of reading
1.2 ... 1.4 bar -5 % of reading
Accuracy of pressure compensation ±0.25 % of reading
Background gas effects for CO
2
and H
2
O, uncompensated
<1 % of reading for <6 vol-% CO
2
<1 % of reading for gas dewpoint < 30 °C
Accuracy of background gas compensation
0 ... 50 vol-% CO
2
±0.5 % of reading
0 ... 300 g/m
3
H
2
O (T
d
= 80 °C) ±1 % of reading
Inputs and Outputs
Power supply
input range 11 ... 36 VDC
Power consumption
maximum 6 W
typical 3 W
Analog output 0/4 ... 20 mA, sourcing
maximum load 500 
Serial output (2-wire, non-isolated) RS-485
Alarm/control relay 30 VAC/60 VDC
Operating Environment
Operating temperature range
for probe (in-line installation) -20 ... +80 °C
for electronics (housing) -40 ... +60 °C
for transmitter (ambient gas measurement) -20 ... +60 °C
Storage temperature range -55 ... +80 °C
Operating pressure range 0.8 ... 1.4 bar
Maximum pressure range for probe up to 10 bar
Compliance
IEC(EN)-61326 Electrical equipment for measurement, control
and laboratory use. EMC requirements
EN50104 Electrical apparatus for the detection and
measurement of oxygen. Performance requirements and test
methods.
EN50271 Electrical Apparatus for the Detection and
Measurement of Combustible Gases, Toxic Gases or Oxygen
Laser safety Class 1 laser product