Chapter 3 _________________________________________________________________Mounting
VAISALA _______________________________________________________________________ 13
Mounting the MMT318
NOTE
Do not unsolder and then again resolder the sensor head cable from
and to the printed board during installation; this procedure may alter
the humidity calibration of the transmitter.
Mounting for Pressurized Pipelines/Oils
Due to the sliding fit, the MMT318 is easy to install into and remove from the
pressurized process. The probe is especially suitable for the measurements in
pipelines.
MMT318 pipe dimensions (in mm); standard 178 (adjustment range 120
mm) and optional 400 mm (adjustment range 340 mm).
Adjust the probe to a suitable depth according to the type of installation and
tighten the clasp nut first manually. Mark the fitting body and the clasp nut.
Tighten the nut a further 50 - 60° (ca.1/6" turn) with a fork spanner. If you have a
suitable torque spanner, tighten the nut to 45±5 Nm (33±4 ft-lbs). Be careful not
to over tighten' the nut.
When the probe is used in a pressurized processes the sensor head should
preferebly be installed through a ball valve assembly.
1. clasp nut, 24 mm hex nut
2. fitting body, 27 mm hex head