Appendix 1 _____________________________________________________________ Applications
VAISALA _______________________________________________________________________ 59
Technical data
Typical measuring
range
0...80* ppm (0...100ºC)
*Upper edge limited to saturation
Accuracy
(Calculation Model with
Avera
e Coefficients
Temperatures > 30 ºC: better than 10 % of the
reading
Tem
eratures < 30ºC: see the fi
ure below.
Figure 2. The maximum errors caused by deviation of mineral oils using
calculation model with average coefficients.
Temperature
Measurement range -40...+180ºC
Response times (with stainless steel filter)
In still air (20ºC) 10 s
In still transformer oil (20ºC) < 10 min
0
2
4
6
8
10
12
14
16
18
20
0 10203040 5060708090100
Tem perature (deg C)
Maximum error of the reading (%)