CAUTION
Take care not to damage the probe body. A damaged body makes
the probe head less tight and may prevent it from going through the
clasp nut.
Use teflon tape or thread sealant to seal the connection between the
fitting body and the process/ ball valve, see Figure below.
Ball Valve Installation Kit for
MMT318
The ball valve installation kit (Vaisala order code: BALLVALVE-1)
is preferred when connecting the probe to a pressurized process or
pipeline. Use the ball valve set or a 1/2" ball valve assembly with a
ball hole of Ø14 mm or more. If you install the sensor head (Ø 12
mm) in a process pipe, please note that the nominal size of the pipe
must be at least 1 inch (2.54 cm). Use the manual press handle to press
the sensor head into the pressurized (< 10 bar) process or pipeline.
1. Shut down the process if the process pressure is more than 10
bars. If the pressure is lower there is no need to shut down the
process.
2. Make the installation according to the figure below. Install the
sensor head transversely against the direction of the process
flow.