Vaisala HUMICAP MMT310 Uživatelská příručka Strana 23

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 81
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 22
Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 23
Ball Valve Installation Kit for MMT318
The optional ball valve installation kit (Vaisala order code:
BALLVALVE-1) is preferred when connecting the probe to a
pressurized process or pipeline. Use the ball valve set or a 1/2" ball valve
assembly with a ball hole of Ø14 mm or more.
If you install the sensor head (Ø 12 mm) in a process pipe, please note
that the nominal size of the pipe must be at least 1 inch (2.54 cm). Use a
manual press handle to press the sensor head into the pressurized (< 10
bar) process or pipeline.
0507-043
Figure 8 MMT318 Ball Valve Assembly Installation
The following numbers refer to Figure 8 above:
1
=
Manual press tool
2
=
Handle of the ball valve
3
=
Probe
4
=
Process chamber or pipeline
5
=
Groove on the probe indicates the upper adjustment limit
6
=
Filter
7
=
Ball of the ball valve
8
=
Fitting screw
Zobrazit stránku 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 80 81

Komentáře k této Příručce

Žádné komentáře