Vaisala HUMICAP MMT310 Uživatelská příručka Strana 44

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 81
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 43
User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M210474EN-C
Select Metric or Nonmetric Output Units
Syntax: UNIT x<cr>
Where
x
=
M or N
M = metric units
N = nonmetric units
Quantity
Metric Unit
Non-Metric Unit
aw water activity
-
aw
T Temperature
°C
T
ppm (for
transformer oil only)
ppm
ppm
Setting, Scaling and Testing the Analog
Outputs
You can select for the two outputs
- current or voltage range (0...20 mA/4...20 mA, 0 … 5/10 V,1 … 5 V
available through scaling) and
- output parameter (aw/ ppm*/ T).
and scale the two outputs according to your needs.
*for transformer oil only
Set Analog Outputs
Syntax: AMODE ch1 ch2<cr>
Where
ch1 and ch2
=
I0 = 0 ... 20 mA
I1 = 4 ... 20 mA
U1 = 0 … 5 V
*
U2 = 0 … 10 V*
*Voltage output applicable from software version 2.0.1 onward.
Example:
>amode i1 u1
Ch1 output mode: 4...20mA
Ch2 output mode: 0...5V
>
Zobrazit stránku 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 80 81

Komentáře k této Příručce

Žádné komentáře