Analog Output Tests
ATEST
Use the serial line to set the analog output test value. The output value
is forced to the inputted value and held until the command is fed without
parameters. The number of quantities as parameters depends on the
number of AOUT-1 modules. The valid range of test voltage/current is
the same as the analog output range.
ATEST [x] [y]
Example:
>atest 1.0
Ch1 : 1.000 (mA/V) H'6644
>
Analog Output Fault Indication Setting
AERR
Use the serial line command AERR to set the analog output error value,
when an error condition is active, in V or mA). The valid range (output
range) for error value depends on the mode of the AOUT-1.
Example:
>aerr
Ch1 error out : 0.000V ?
>
where
x = volts
y = milliamperes