General Settings
Changing Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 1 on page 18 for available quantities.
MPCP1/
MPCP2/
MPCP3
Performs a multipoint correction for the
barometer module/module.
CTEXT Shows or sets the calibration info text.
NOTE
The calibration and adjustment commands are available in adjustment
mode only. Press the adjustment button before inputting these
commands.
Setting and testing the analog outputs
AMODE Displays the analog output mode (if an
AOUT-1 module/s is/are connected).
ASEL Sets the analog output quantity and
scaling (low/high).
ACAL Adjusts the analog output.
AERR Sets the analog output error value.
ATEST Sets the analog output test value
Setting and testing the relays
RSEL Sets the relay scaling (if a RELAY-1
module/s is/are connected).
RTEST Sets a test value for a relay output.
NOTE
Only the quantities selected when ordering the device can be selected
as a display output quantity.