User's Guide ______________________________________________________________________
80 __________________________________________________________________ M210855EN-C
DPMAX [x]
Example:
>dpmax 0.2
Max. diff. : 0.20 hPa
>
Use the AVRG command to set barometer measurement averaging
time (in seconds). Valid range of averaging time is 1...600 seconds.
AVRG [x]
Example:
>avrg
Average filter : 1.0 s ?
>
HHCP
The HHCP command shows or sets the altitude for height corrected
pressure. The valid range of HCP height is -30...+30 m.
HQFE
The HQFE command displays or sets altitude for QFE corrected
pressure. The valid range of QFE height is -30...+30 m.
HQNH
The HQNH command displays or sets altitude for QNH corrected
pressure. The valid range of QNH height is -30...3000 m.
where
x = Pressure
where
x = Average time