Vaisala DEWCAP DM500 Uživatelská příručka Strana 31

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 138
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 30
Chapter 4 _________________________________ General Information About Sampling and Results
VAISALA _______________________________________________________________________ 29
The Effect of the Sample Gas Flow Rate on
Cooling Capacity and Result Deviation
The cooling effect of the gas flow rate is
shown in Figure 9 (on the right). Low flow
rate increases cooling capacity but also
increases response time.
-79
-78
-77
-76
-75
-74
-73
-72
0 0.5 1 1.5 2 2.5
Flow l/min
Tsensor /C
Figure 9 Effect of Sample Flow Rate on Lowest
Achievable Sensor Temperature
The sample flow rate has an effect on the
deviation of the measurement results as
shown in Figure 10 (on the right).
Sample flow rate range: 0.5...1.5 slpm.
Recommended sample flow rate: 1.0 slpm.
-0.2
-0.15
-0.1
-0.05
0
0.05
0.1
0.15
0.2
0.5 0.75 1 1.25 1.5
flow/slpm
deviation/°C
Figure 10 Effect of Sample Flow Rate on
Dewpoint Measurement at -60 ºC
Zobrazit stránku 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 137 138

Komentáře k této Příručce

Žádné komentáře