Chapter 16 _____________________________________________________ Serial Line Commands
VAISALA _______________________________________________________________________ 87
command R does not change the default operation mode set with the
command SMODE.
Example:
>r
0.000 'C 0.00 'C 0.00 hPa 0.00
slpm
0.000 'C 0.00 'C 0.00 hPa 0.00
slpm
S Stopping the Continuous Output
Syntax: S<cr>
Type S to stop outputting.
ADDR Setting the Device Address
Syntax: ADDR xx<cr>
where
xx = 0...99 (address)
The address is used in POLL mode when more than one device is
connected to one serial line.
SEND Outputting One Single Reading
Syntax: SEND x<cr>
where
x = device address (0...99) in POLL state
INTV Setting the Output Interval
Syntax: INTV xxx yyy<cr>
where
xxx = output interval
yyy = unit (s, min, h)