User's Guide ______________________________________________________________________
62 __________________________________________________________________ M210205EN-D
Response Speed
You can set the response speed to stable, normal (default) or fast. If
fast transients in the dewpoint prevent the system from reaching
balance, the signal can be stabilized by selecting the Stable response
speed. Fastest response time is achieved by setting the speed to Fast
but this may increase the noise in the dewpoint measurement. As a
rule of thumb, always use the fastest setting that gives stable readings!
1. Open MENU: press .
2. Select Measuring, press .
3. Select
Measurement settings, press .
4. Select
Response, press SET.
5. Select one of the following options:
Stable: Press SELECT to have the slowest response speed.
Normal: Press SELECT to have normal response speed.
Fast: Press SELECT to have fast response speed.
6. Press EXIT to return to the main view.
Pump ON/OFF
1. Open MENU: press .
2. Select
Measuring, press .
3. Select Measurement settings, press .
4. Select Pump, press ON/OFF.
5. Press EXIT to return to the main view.