Chapter 14 _________________________________________________________ Output Interfaces
VAISALA _______________________________________________________________________ 75
3. Select Communications, press .
4. Select Comm. mode, press SET.
5. Select one of the following mode, press SELECT:
STOP: Outputs measurements only by command. All
commands can be used.
RUN: Outputs measurements automatically in a format set by a
serial command FORM. Only serial command S can be used.
POLL: Outputs measurements only with the serial command
SEND.
MI70 Link: Sets the MI70 Link program communication
parameters. Other communication settings have no effect when
using MI70 Link software.
6. Set the polling address if needed, select Polling address, press
SET.
7. Give the address for POLL communication mode by using
numerical buttons, press OK.
8. Select Baud rate, press SET.
9. Select one of the following baud rate, press SELECT:
300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600,115200
bps
10. Select Serial format, press SET.
11. Select the data bits (7 or 8), parity (N = none, E = even,
O = odd) and stop bits (1 or 2), press SELECT.
12. Select Echo or Half duplex, press NO/YES to turn the functions
on/off.
13. To return to the main view, press
EXIT.
Analog Outputs (Optional)
DMI500 user interface can be equipped with three optional 4...20 mA
passive current outputs. These 2-wire channels must be powered with
12...28 VDC external power supplies (not provided). The analog
outputs are galvanically isolated from electronics but they are not
isolated from each other. Thus, you need a separate (isolated) power
supply for each monitored analog channel.