User's Guide _______________________________________________________________________
58 ___________________________________________________________________ M210934EN-D
Calibration and Adjustment Using MM70 and a
Reference Environment
Required equipment:
- A fully charged MI70 indicator (the measurement display that is
included in the MM70 package)
- Connection cable for MM70 hand-held meter (219980)
- One or two reference environments that fulfill the requirements listed
in section Adjustment Points on page 55.
For example, you can use the Vaisala HMK15 humidity calibrator for a
two-point humidity calibration and adjustment. Use the LiCl (11 %RH)
and NaCl (75 %RH) salt solutions. You also need the special cover set
for HMK15 (230914). For two-point calibrations using HMK15, use a
dedicated cover for each salt chamber to avoid long stabilization times
and possible contamination of the salt references.
1. Plug in the connection cable 219980 to port II of the MMT162, and
the other end to port II of the MI70 indicator.
2. If any cable is connected to port I of the MMT162, disconnect it.
3. If the MMP70 probe is connected to the MI70 indicator, disconnect
it.
4. Remove the filter from the MMT162 and insert it in the reference
environment (for example, the LiCl (11 %RH) salt chamber).
5. Wait for 30 minutes for temperature and humidity to stabilize.
6. Turn on the MI70 indicator.
7. Select Functions Adjustments from the menu of the MI70 and
press Start.
Zobrazit stránku 59