Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA ________________________________________________________________________ 59
8. The adjustment mode starts. Automatic power off is disabled while
in adjustmend mode; press OK to acknowledge the message.
9. Select the parameter to calibrate and adjust (RH or T) from the list
and press Select. This procedure is applicable for both parameters,
but the following steps will assume you selected RH.
10. The adjustment mode should now show a stable, unchanging RH
value. The difference between the measured value and the
reference humidity should be no greater than the combined
measurement uncertainty of the MMT162 transmitter and the
reference environment. Depending on the result, proceed as
follows:
- If the MMT162 transmitter is within its accuracy
specification, there is no need to proceed with the adjustment.
Select Back and Exit to leave the adjustment mode.
- If adjustment is required, continue with the next step.
11. Press Adjust and select 1-point adjustment or 2-point
adjustment.
12. Press Ready and enter the correct value (the reference value) for
the parameter you are adjusting. Press OK when done, and YES
when asked if you really want to adjust.
13. You have now entered the first adjustment point. If you selected 1-
point adjustment, you will see the text Adjustment done. You
have now completed the 1-point adjustment. Select Back and Exit
to leave the adjustment mode.
If you are doing a 2-point adjustment, the indicator shows the
screen for the second reference point. Continue with the next step.
If something went wrong with the adjustment, check that your
reference environment fulfills the requirements listed in section
Adjustment Points on page 55, and that the measurement had
properly stabilized. Retry the adjustment after checking.
14. Transfer the MMT162 to the second reference environment (for
example, the NaCl (75 %RH) salt chamber).