APPENDIX 7 ________________________________________________________________________
136 ____________________________________________________________________ U263EN-1.4
Setting and activating the outputs using an RS line
Disconnect the security lock jumper!
ALARM <cr>
or
ALARM n ON <cr>
where n = channel number (1 or 2)
The currently valid settings of the alarm output unit can be checked
with command ALARM:
>alarm<cr>
Ch1 RH LO OFF 0.00 0.00 %RH
Ch2 RH HI OFF 100.00 0.00 %RH
>
This is an example of the factory setting (both alarm outputs are in
OFF mode, i.e. not in use). The settings can be changed with
command ALARM:
>alarm 1 Td HI -20 5
Ch1 Td HI OFF -20.00 5.00 ’C
Ch2 RH HI OFF 100.00 0.00 %RH
>
When giving this command, first enter the channel number i.e. the
number of the alarm output you wish to use (1 or 2). Then select the
quantity you wish to have on that channel (Td, ppm, RH or T). The
third parameter (HI/LO) determines, whether the alarm is activated
when the setpoint value is exceeded (HI) or not reached (LO). The
fourth parameter is the actual setpoint value, which activates the
alarm. The last parameter is the hysteresis value; it indicates how
much the measured value has to exceed or go below the setpoint
before the alarm is deactivated. If the third parameter is HI, the alarm
is deactivated when the measured value goes below the setpoint with
the chosen hysteresis value. If the parameter is LO, the alarm is
deactivated when the measured value exceeds the setpoint with the
chosen hysteresis value.