The 0D[G7GS parameter defines the maximum allowed change of the
dewpoint value during the pre-defined G7GS time. If the change in
dewpoint value exceeds the limit, the calibration is not started. The
auto-calibration is done only after process is stabilized. The
transmitter retries untill the calibration is succesfully complited. Note
that normally, this parameter does not have to be changed.
The G7GSWLPH defines the measuring time of the change in dewpoint
prior to the auto-calibration (see the parameter 0D[G7GS.
The Max corr. parameter defines the maximum correction (%RH) the
transmitter does during each auto-calibration cycle.
The 6HWWO WLPH defines the time the output values prior to the auto-
calibration are frozen after the calibration. The time is for sensor
temperature stabilization.
NOTE
Normally, the Max dTdp and the Max corr. parameters need not be
changed.
NOTE
Auto-calibration takes place only when the security lock jumper is
connected. With command <fst on>, the serial line printing indicates
the stages of the auto-calibration (n = no auto-calibration, w,h,s =
different phases in auto-calibration, S = settling time auto-calibration
is on).
DCAL<cr>
>dcal
Calibration...
any key to abort
> (appears when the auto-calibration
is completed in 60 - 70 seconds)
If you press any key, the calibration is interrupted. If no key is
pressed, the calibration takes place immediately. If the process is
unstable or relative humidity is over 10%, the prompt (>) appears
immediately. In that case the calibration is not performed. The
calibration is not performed again even if the maximum correction is
exceeded. (see Chapter Display/keypad commands)