APPENDIX 1 SERIAL COMMANDS __________________________________________________________
90 _____________________________________________________________________ U263EN-1.4
S Stopping the measurement output
S<cr>
Ends the RUN state; after this command all other commands can be used.
SEND Outputting a reading once
SEND <cr> in STOP state
or
SEND aa <cr> in POLL state
aa = address of the transmitter when more than one transmitter is
connected to a serial bus (0...99; set with command ADDR)
Outputs the current measurement readings via the serial line. The output type is
the following:
Td= -9.3 ’C PPM= 2733 T= 22.1 ’C RH= 10.4 %RH
Td= -9.3 ’C PPM= 2730 T= 22.1 ’C RH= 10.4 %RH
...
The output format can be changed with command FORM.
DSEND Outputting readings of all connected transmitters
once
DSEND <cr>
All transmitters connected to the serial bus send their addresses and current
measurement readings in an order defined by their addresses. After receiving
DSEND command a transmitter sets a delay time according to its address value
and sends the data after this delay. DSEND works also in POLL mode. With
this command, the user can for example easily find out the addresses of the
transmitters.
The output when four transmitters with addresses 4, 5, 10, 33 have been con-
nected to the serial bus:
>dsend <cr>
4 14.43 %RH
5 22.7 ’C
10 14.99 %RH
33 22.3 ’C
>