Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 105
Selecting Data Recording Quantities
If the device is provided with the optional display, the recorded quantities
are always those selected for the display. Up to three quantities can be
recorded at a time. For instructions on how to select the display quantities
with the keypad, see section Changing Quantities and Units on page 92.
DSEL
Use the serial line command DSEL to select the quantities to be recorded
if the transmitter is not equipped with display/keypad.
DSEL [xxx]
where
xxx = Data recording quantity. See Table 3 on page 19 and Table 4
on page 19 for the quantities.
Example:
>dsel rh t tdf
RH T Tdf
>
Enter the command without parameters and press ENTER to display the
current recording parameters.
View Recorded Data
If the device is provided with the optional display, the graphical display
shows the data of the selected quantities, one at a time. See section
Graphic History on page 66 for details about graphical display.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DIR
Use the serial line and enter the DIR command to check the available
files.
Without the data logger module, the device records six files (six
observation periods) for each selected quantity. The data logger raises the
number of recorded files to seven for each quantity. Thus, the total
amount of the files varies between 6 and 21. See Table 11 on page 67.
Select, for example, three quantities (RH, T, and Tdf). The last column
illustrates the number of data points that has been stored in the file.