Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 45
0503-033
Figure 27 Cleaning of Tightening Cone
The following numbers refer to Figure 27:
1 = Fitting screw
2 = Sealing washer
3 = Tightening cone
4 = Clean cotton stick
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
NOTE
When installed in a process with a pressure differing from normal
atmospheric pressure, please enter the pressure value of the process (in
hPa or mbar) into the transmitter memory via the serial line (see
command PRES and XPRES on page 96) or by using display/keypad.
HMT335 for High Temperatures
HMT335 is installed similarly than the HMT333 probe but without the
supporting bar. Refer to Appendix A on page 163 for more information
on the duct installation kit for HMT335.
To avoid incorrect humidity readings, the tem
perature differences
between inside and outside of the duct must not be remarkable.