Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 121
Example:
>sdelay
Serial delay : 0 ? 10
>sdelay
Serial delay : 10 ?
SERI
Use the SERI command to input RS-485 bus settings.
SERI [b p d s]
where
b
= bit rate (300, 600, 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200)
= parity (n = none, e = even, o = odd)
d
= data bits (7 or 8)
s
= stop bits (1 or 2)
ECHO
Use the ECHO command to enable/disable echo of characters received
over the serial line.
ECHO [x]
where
x = ON/OFF (default = OFF)
When using 2-wire connection, echo must be always disabled.
SMODE
Use the SMODE command to set the default serial interface mode.
SMODE [xxxx]
where
xxxx
= STOP, RUN or POLL
In STOP mode: measurements output only by command SEND, all
commands can be used
In RUN mode: outputting automatically, only command S can be
used to stop.
In POLL mode: measurements output only with command SEND
[addr].