User's Guide ______________________________________________________________________
92 __________________________________________________________________ M210566EN-H
g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa
h= 43.5 kJ/kg
>
For more information on chemical purge, see section Chemical Purge
(Optional) on page 124.
Where the state of the probe is indicated by the following letters and values:
... xxx =
ormal operation where h xxx= Probe heat power
X ... xxx = Sensor heating where xxx = Sensor temperature (ºC)
H ... xxx = Chemical purge where xxx = Sensor temperature (ºC)
S ... xxx = Sensor cooling
after purge
where xxx = Sensor temperature (ºC)
General Settings
Changing Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 3 on page 19 for available quantities and
Table 4 on page 19 for optional quantities.
NOTE
Only the quantities selected when ordering the device can be selected as
a display output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Press ► arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display
quantities at a time.
5. Press EXIT to return to the basic display.
To select display units:
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Use the ▲▼arrow buttons to select Units. Confirm the selection
by pressing the right-hand arrow button.