User's Guide ______________________________________________________________________
126 _________________________________________________________________ M210566EN-H
Starting and Configuring Chemical Purge
Using Buttons on Motherboard
Start manual chemical purge by pressing simultaneously two PURGE
buttons on the motherboard inside the transmitter for a few seconds.
Indicator led flashes until purge is complete (up to 6 minutes).
0508-011
Figure 65 Purge Buttons on Motherboard
Using Display/Keypad (Optional)
Set the automatic and manual chemical purge by using the
display/keypad.
1. Open the MAIN MENU by pressing any of the ▼▲◄► arrow buttons.
2. Select Measuring, press button.
3. Select Chemical purge, press button.
0706-006
Figure 66 Chemical Purge Settings
- Turn on/off the automatic purge by selecting Automatic purge,
press ON/OFF button.
- Set the automatic purge interval by selecting Interval: ..., press
SET. Set the purge interval and the unit (hour/day) by using the
arrow buttons. The interval must be 1 hour ... 10 days. Press OK.
- Select Start-up purge by using the arrow buttons. Press On/Off to
turn the start-up purge on/off.
- Start manual purge by selecting Manual purge and pressing
START.